Based on the claim against the product by the customer noting an issue with an insturment, an investigation was completed to determine the cause of this reported event.The large hem-o-lok clip applier instrument was analyzed and found to have failed the clip test due to misapplication of the clip.The instrument passed engagement and was able to retain the clip, but it could not apply the clip.The instrument was found to have a loose grip cable, which caused it to fail to apply the clip to the test tubing on multiple attempts.Additional observation(s) not reported by the site: the instrument was found to have a loose grip cable at the distal end.The complaint regarding issue with the instrument was confirmed by failure analysis, which indicates that the device did contribute to the customer reported issue.
|
It was reported that during a da vinci-assisted surgical procedure, the large hem-o-lok clip applier instrument was broken.There is no further information available.Intuitive surgical, inc.(isi) followed up with the initial reporter; however, the customer could not provide any additional information.
|