The sample was received for evaluation with a patient connector and shield guard attached to the female connector.A visual inspection with the naked eye noted a separation between the female connector and main body.Functional testing including clear passage testing and clamp function testing were performed with no issues noted.Leak testing failed due to a leak from a cut/hole in the silicone tubing.A connection and disconnection between the returned patient connector and the female connector were performed by hand with no issues or leaks noted.The reported issue of unable to disconnect was not duplicated.The cause of the separation was determined to be manufacturing related due to an inadequate solvent bond between the female connector, insert chip, and main body.The cause of the cut in tubing was undetermined; however, the cut most likely occurred during the disconnection attempt.Should additional relevant information become available, a supplemental report will be submitted.
|