Model Number 799-EMVP-01-01 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 08/09/2023 |
Event Type
malfunction
|
Event Description
|
Complainant alleged that while attempting to treat a 59-year-old male patient, the device displayed an "internal comm failure - 1475" error message.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including stress testing, complete calibration, exhalation backup and autocal valve testing without duplicating the report.A ribbon cable was replaced as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|