The device was returned to olympus for evaluation.The device evaluation found the complaint was confirmed and the bending section adhesive was chipped.Also, evaluation found the following: due to damage on forceps elevator lever (surgery product), the angle knob torque of forceps elevator lever at engage exceeds the standard value; bending section cover or distal sheath rubber has a scratch; and bending section cover or distal sheath rubber has a chip.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the reportable event occurred due to stress of repeated use, external factors, or handling of the device.The event can be prevented by following the instructions for use (ifu) which state: "chapter 3 preparation and inspection, section 3.8 inspection of the endoscopic system ¦ inspection of the endoscopic image confirm that the wli (white light imaging) and nbi (narrow band imaging) endoscopic images are normal.1.Before inspection, wipe the objective lens using clean lint-free cloths moistened with saline solution or sterilized water.2.Observe the palm of your hand in the wli and nbi endoscopic images.3.Confirm that light is output from the endoscope¿s distal end.4.Adjust the brightness level as appropriate.5.Confirm that the wli and nbi endoscopic images are free from noise, blur, fog, or other irregularities.6.Turn the angulation control levers slowly in each direction until it stops.7.Confirm that the wli and nbi endoscopic images do not momentarily disappear or display any other irregularities." olympus will continue to monitor field performance for this device.
|