This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation the bulge in the channel was confirmed.However, the specific cause of the bulge was not established at this time.¿do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.Never perform angulation control forcibly or abruptly.Never forcefully pull, twist, or rotate the angulated bending section.Patient injury, bleeding, and/or perforation may result.It may also become impossible to straighten the bending section during an examination.¿ olympus will continue to monitor field performance for this device.
|