Catalog Number 221261 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/23/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that prior to using bd bbl¿ trypticase soy agar with 5% sheep blood (tsa ii), two sleeves were received with growth on them.No patient impact reported.The following information was provided by the initial reporter: "customer reports receiving plates with growth on them.".
|
|
Manufacturer Narrative
|
D.4.Medical device lot #: unknown.D.4.Medical device expiration date: unknown.H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.H.4.Device manufacture date: unknown.
|
|
Event Description
|
It was reported that prior to using bd bbl¿ trypticase soy agar with 5% sheep blood (tsa ii), two sleeves were received with growth on them.No patient impact reported.The following information was provided by the initial reporter: "customer reports receiving plates with growth on them.".
|
|
Manufacturer Narrative
|
H.6.Investigation summary: two photos were received for investigation.One photo shows the side of a sleeve with growth visible in the top plate.The other photo shows the side of a sleeve with one darker color from contamination highlighted.No return samples were received for investigation.No batch number was provided for investigation.Retention samples cannot be evaluated without a specific batch number.The appropriate quality databases were reviewed, and bd identified a contamination trend for this product that would include batches still within expiry at the time this complaint was taken.The investigation found opportunities for bioburden reduction in the manufacturing process.A capa (corrective and preventative actions) has been initiated and involves implementing additional cleaning events and evaluation of manufacturing procedures focused on in-process bioburden reduction.Additional trainings are part of the implementation with an ongoing training review for cleaning processes.Improvement in observation of contamination is expected.This complaint can be confirmed based on the trend.Bd will continue to trend complaints for contamination.
|
|
Search Alerts/Recalls
|