MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II); CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II); CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL Back to Search Results
Catalog Number 221261
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2023
Event Type  malfunction  
Event Description
It was reported that prior to using bd bbl¿ trypticase soy agar with 5% sheep blood (tsa ii), two sleeves were received with growth on them.No patient impact reported.The following information was provided by the initial reporter: "customer reports receiving plates with growth on them.".
 
Manufacturer Narrative
D.4.Medical device lot #: unknown.D.4.Medical device expiration date: unknown.H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.H.4.Device manufacture date: unknown.
 
Event Description
It was reported that prior to using bd bbl¿ trypticase soy agar with 5% sheep blood (tsa ii), two sleeves were received with growth on them.No patient impact reported.The following information was provided by the initial reporter: "customer reports receiving plates with growth on them.".
 
Manufacturer Narrative
H.6.Investigation summary: two photos were received for investigation.One photo shows the side of a sleeve with growth visible in the top plate.The other photo shows the side of a sleeve with one darker color from contamination highlighted.No return samples were received for investigation.No batch number was provided for investigation.Retention samples cannot be evaluated without a specific batch number.The appropriate quality databases were reviewed, and bd identified a contamination trend for this product that would include batches still within expiry at the time this complaint was taken.The investigation found opportunities for bioburden reduction in the manufacturing process.A capa (corrective and preventative actions) has been initiated and involves implementing additional cleaning events and evaluation of manufacturing procedures focused on in-process bioburden reduction.Additional trainings are part of the implementation with an ongoing training review for cleaning processes.Improvement in observation of contamination is expected.This complaint can be confirmed based on the trend.Bd will continue to trend complaints for contamination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II)
Type of Device
CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17653291
MDR Text Key322318298
Report Number1119779-2023-00958
Device Sequence Number1
Product Code JSG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PREAMENDMENT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number221261
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/23/2023
Initial Date FDA Received08/30/2023
Supplement Dates Manufacturer Received03/12/2024
Supplement Dates FDA Received03/25/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-