Device evaluation by manufacturer: the returned product to boston scientific consisted of a jetstream xc-2.1 atherectomy catheter.The device was visually and microscopically examined for any shaft damage.Visual examination showed a kink on the catheter shaft located 1cm from the tip and a severe kink located 69.5cm from the tip.The device was set-up and functionally tested per the instructions for use.The device primed and ran as designed.It was noticed that the catheter shaft leaked fluid from the kink at the 69.5cm location.Testing the device in the anatomical was not possible due to the damage on the catheter shaft.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
Reportable based on device analysis completed on 17aug2023.It was reported that difficulty advancing occurred.A 2.1mm jetstream xc atherectomy catheter was selected for the atherectomy procedure.During the advancement over the non-boston scientific wire, however, the jetstream catheter was not able to move forward.The rex function was possible, and the catheter was removed intact.Upon removal, there was a visible bend directly behind the blades of the catheter.The device was replaced with a new jetstream catheter to complete the procedure.No patient complications were reported.However, device analysis revealed the infusion line was leaking.
|