BAYLIS MEDICAL COMPANY INC. VERSACROSS CONNECT LAAC ACCESS SOLUTION; DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION Back to Search Results
Lot Number VMFE100423
Device Problems Difficult to Remove (1528); Defective Device (2588); Device-Device Incompatibility (2919); Material Deformation (2976); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2023
Event Type  malfunction  
Event Description
It was reported that a versacross connect access solution was selected for use during a left atrium appendage closure.During preparation, it was noted that the mechanical guidewire was not coming out of the dilator (about an inch exposed from dilator), once trying to exchange the mechanical guidewire for the versacross rf wire.It was not a smooth transition among the devices.Thus, it was needed to remove the dilator and mechanical guidewire together, hence force had to be applied to remove the guidewire from the dilator, when outside the patient anatomy.The mechanical guidewire was noted kinked and unraveled, and in one piece.The procedure was continued along with a versacross rf wire and the same dilator, and it was completed successfully.No patient complications were reported.The device is not expected to be returned for analysis.
 
Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACROSS CONNECT LAAC ACCESS SOLUTION
Type of Device
DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION
Manufacturer (Section D)
BAYLIS MEDICAL COMPANY INC.
5959 trans-canada highway
montreal H4T 1 A1
CA  H4T 1A1
Manufacturer (Section G)
BAYLIS MEDICAL COMPANY INC.
5825 explorer drive
mississauga, on L4W 5 P6
CA   L4W 5P6
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key17670813
MDR Text Key322510476
Report Number2124215-2023-47280
Device Sequence Number1
Product Code DRE
UDI-Device Identifier00685447012597
UDI-Public00685447012597
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K220414
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot NumberVMFE100423
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/10/2023
Initial Date FDA Received09/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/10/2023
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age81 YR
-
-