This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the reported event could not be identified.However, it¿s likely the cause is related to a previous third-party repair in which a different type of adhesive was used.Olympus does not recommend having repairs carried out by non-authorized service centers, as the inner shaft can no longer be of standard quality and, in the worst case, may endanger patients and staff.The suggested event is inspectable and preventable by handling the device in accordance with the following sections of the instructions for use: 4 before use warning infection control risk properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing inspecting the product visually inspect the product.Make sure that it has: -- no corrosion -- no dents -- no scratches ceramic insulation at distal end visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning risk of injury impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.Olympus will continue to monitor field performance for this device.
|