Model Number 60-05-65
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/08/2023
Event Type  malfunction  
Event Description
Livanova deutschland has received a repor that the elctronic remote clamp had an errore during a procedure.There is no report of any patient injury.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 erc tubing clamp.The incident occurred in united states.A livanova filed service engineer was dispatched to the customer site.The complained erc engage/close on its own not reproducible.The device was replaced with another one.Functional checks were successfully completed.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
The unit has been installed since 2014 and an analysis of the complaints database did not reveal further reported incidents.The most likely root cause of the reported issue are: defective internal board resulting in an intermittent failure or faulty communication between the mentioned erc and the related s5 console.The wearing of electro-mechanical device can be originated by the specific use condition at customer site and may have contributed to the event, the risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See intial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCP WITH TUBING CLAMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17681065
MDR Text Key322632362
Report Number9611109-2023-00415
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-05-65
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/08/2023
Initial Date FDA Received09/05/2023
Supplement Dates Manufacturer Received10/24/2023
Supplement Dates FDA Received11/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-