Model Number 60-05-65 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/08/2023 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland has received a repor that the elctronic remote clamp had an errore during a procedure.There is no report of any patient injury.
|
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 erc tubing clamp.The incident occurred in united states.A livanova filed service engineer was dispatched to the customer site.The complained erc engage/close on its own not reproducible.The device was replaced with another one.Functional checks were successfully completed.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
The unit has been installed since 2014 and an analysis of the complaints database did not reveal further reported incidents.The most likely root cause of the reported issue are: defective internal board resulting in an intermittent failure or faulty communication between the mentioned erc and the related s5 console.The wearing of electro-mechanical device can be originated by the specific use condition at customer site and may have contributed to the event, the risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
|
Event Description
|
See intial report.
|
|