MEDTRONIC, INC. ACHIEVE ADVANCE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT Back to Search Results
Model Number 2ACH20
Device Problems Material Deformation (2976); Material Twisted/Bent (2981); Device Handling Problem (3265)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/21/2023
Event Type  malfunction  
Manufacturer Narrative
Select patient information cannot be provided due to regional privacy regulations.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, the front end of the mapping catheter was kinked after it was removed from the packaging and was unable to bend back into shape.It was also reported that the mapping catheter was unable to support the exploration of pulmonary veins and balloon occlusion during surgery and it was soft in comparison to the earlier model.The mapping catheter was replaced which resolved the issue. the case was completed with cryo. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the 2ach20 mapping catheter with lot 8183397 was returned and analyzed.Visual inspection was performed, and a kink/twist was observed on the catheter shaft.Additionally, a broken shaft was observed.Visual inspection of the loop segment area showed the loop was intact with no apparent issues.No damage was observed along with the tip/loop section of the mapping catheter.Visual inspection of the pebax tubing area showed the pebax tubing was intact with no apparent issues.No damage was observed along with the pebax tubing section of the mapping catheter.Visual inspection of the electrode(s) showed the electrode(s) were intact with no apparent issues.All electrodes exist on the loop section and no cosmetic issue or anomalies were identified.Visual inspection of the shaft segment area showed the shaft was broken approximately 57.25 inches from the lemo connector.No broken electrocardiogram (ecg) signal wires were observed.Visual inspection of the shaft segment area showed the shaft was kinked approximately 48.5 inches from the lemo connector.In conclusion, the reported mapping catheter kink was confirmed.Additionally, the mapping catheter failed the return product inspection due a broken shaft.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE ADVANCE MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17683034
MDR Text Key322634826
Report Number2182208-2023-02474
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K162892
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2ACH20
Device Catalogue Number2ACH20
Device Lot Number8183397
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/21/2023
Initial Date FDA Received09/05/2023
Supplement Dates Manufacturer Received09/18/2023
Supplement Dates FDA Received10/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age64 YR
Patient SexMale
Patient Weight60 KG
-
-