• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 III; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 III; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 07976887190
Device Problems Low Test Results (2458); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/07/2023
Event Type  malfunction  
Manufacturer Narrative
The analyzer serial number is (b)(6).The investigation is ongoing.
 
Event Description
There was an allegation of questionable elecsys ft4 iii assay results for 1 patient sample on a cobas e 801 module.The initial ft4 result was 0.838 ng/dl.The repeat result was 1.04 ng/dl.
 
Manufacturer Narrative
The calibration and qc data provided was acceptable.Based on the calibration and qc data, a general reagent issue could be excluded.The investigation did not identify a product problem.The root cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17684689
MDR Text Key322674532
Report Number1823260-2023-02885
Device Sequence Number1
Product Code CEC
Combination Product (y/n)Y
Reporter Country CodeCS
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07976887190
Device Lot Number70021600
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2023
Initial Date FDA Received09/05/2023
Supplement Dates Manufacturer Received09/07/2023
Supplement Dates FDA Received09/28/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age70 YR
Patient SexFemale
-
-