ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE DR; Pulse generator, permanent, implantable Back to Search Results
Model Number CD2411-36Q
Device Problems Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574); Improper or Incorrect Procedure or Method (2017); Low impedance (2285); Charging Problem (2892)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2023
Event Type  Injury  
Event Description
Related manufacturer report number: 2017865-2023-41076 it was reported during remote follow up that the implantable cardioverter defibrillator and its associated right ventricular (rv) lead showed evidence of low shock impedance and an over current detection alert.The device was also found to have delivered inappropriate shock due to incorrect interpretation of signal.During in clinic interrogation, it was discovered that the device had been mis-programmed with settings that were not correct for the rv lead.The device was successfully reprogrammed to the appropriate setting.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE DR
Type of Device
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17686499
MDR Text Key322695176
Report Number2017865-2023-41079
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507615
UDI-Public05414734507615
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Model NumberCD2411-36Q
Device Lot NumberA000048971
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/16/2023
Initial Date FDA Received09/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/09/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age52 YR
Patient SexMale
Patient Weight95 KG
-
-