BOSTON SCIENTIFIC NEUROMODULATION VERCISE GENUS; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
|
Back to Search Results |
|
Model Number DB-1216 |
Device Problems
Unexpected Therapeutic Results (1631); Power Problem (3010); Wireless Communication Problem (3283)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/11/2023 |
Event Type
Injury
|
Event Description
|
It was reported that following a non-device related procedure, the patient experienced a loss of stimulation, and the patients deep brain stimulation (dbs) implantable pulse generator (ipg) would not turn on.In addition, the ipg would no longer connect with the clinician programmer.The patient underwent a revision procedure in which the ipg was replaced.The procedure was completed successfully and there were no patient complications.
|
|
Event Description
|
It was reported that following a non-device related procedure, the patient experienced a loss of stimulation, and the patients deep brain stimulation (dbs) implantable pulse generator (ipg) would not turn on.In addition, the ipg would no longer connect with the clinician programmer.The patient underwent a revision procedure in which the ipg was replaced.The procedure was completed successfully and there were no patient complications.
|
|
Manufacturer Narrative
|
Db-1216, serial (b)(6): the returned ipg passed visual inspection.However, it was unable to be recharged after three four-hour charge cycles.It was cut open and the device exhibited high sleep current.Electrical testing revealed a low vh resistance measurement, which indicates an electrical short within the application-specific integrated circuit (asic).The damaged asic resulted in the premature battery depletion post non-device related procedure.This type of damage is typically caused when the ipg is exposed to high voltage transients, high current sources, and high radio frequency (rf) energy.Therefore, the ipg was likely damaged during the non-device related procedure.Additionally, a review of the instructions for use (ifu) revealed that medical therapies or procedures such as electrocautery may turn stimulation off or may cause permanent damage to the stimulator, particularly if used in close proximity to the device.The ifu further states that electrocautery can transfer destructive current into the dbs leads and-or stimulator.Electrocautery probes must be kept a minimum of 1 inch away from the implanted device.
|
|
Search Alerts/Recalls
|
|
|