SYNTHES GMBH 2.8MM DRILL BIT/QC/165MM Back to Search Results
Catalog Number 310.284
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2022
Event Type  malfunction  
Event Description
Device report from colombia reports an event as follows: it was reported that during a procedure on (b)(6) 2022, two drill bits had to be changed out due to defects.Procedure was completed with no delay.There was no patient consequence.Upon manufacturer investigation, it was determined that one drill bit had the cutting edges of the fluted tip blunt and worn.This report is for a 2.8mm drill bit/qc/165mm.This is report 2 of 2 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h6: a manufacturing record evaluation was performed for the finished device.Product code: 310.284 lot: 93p3909 it was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: march 3, 2021.Manufacturing site: jabil bettlach.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that lcp drill bit ø2.8 w/stop l165 2flute f/ had the cutting edges of the fluted tip blunt and worn.The observed condition was consistent as an end of life indicator for the device.No other issues were identified.The device exhibits damage consistent with repeated use.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed blunt and worn condition of lcp drill bit ø2.8 w/stop l165 2flute f/ would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: drill bit 2-flute dx.X with stop ring (current and manufactured) device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.8MM DRILL BIT/QC/165MM
Type of Device
DRILL BIT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17689685
MDR Text Key322714876
Report Number8030965-2023-11163
Device Sequence Number1
Product Code HTW
UDI-Device Identifier07611819245891
UDI-Public(01)07611819245891
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number310.284
Device Lot Number93P3909
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/24/2023
Initial Date FDA Received09/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/03/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
DRILL BIT Ø2.8 W/SCAL L200/100 3FLUTE F/
-
-