SHIRAKAWA OLYMPUS CO., LTD. OES HYSTEROFIBERSCOPE Back to Search Results
Model Number HYF-XP
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, during reprocessing, the oes hysterofiberscope tested positive on (b)(6) 2023, for 2 colony forming units (cfus) of an unspecified nonpathogenic microorganism, it was not specified which channels were sampled.The user did not report any contamination or any other serious deterioration in the state of health of any person, to which the scope could have been a contributory cause.
 
Manufacturer Narrative
The suspect device has not yet been returned to olympus for evaluation, therefore the physical device evaluation has not been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis results are pending.The customer provided the cleaning, disinfection, and sterilization (cds) processes performed at the user facility.The customer did confirm that there were no deviations or deficiencies concerning reprocessing of the scope.Additionally, the customer confirmed that there were no suspected patient infections due to the facility findings.The customer did not provide the specific steps taken during the cleaning sterilization and disinfection (cds) process.The investigation is ongoing, a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information provided by the third party microbiological results: updated field: h10 the hygiene microbiological investigation report indicated the channels of the scope were cultured and found <1 cfu of an unspecified microorganism.The results obtained comply with the target level for an endoscope subjected to high level disinfection and rinsed with sterile water.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The device was evaluated where no abnormalities were found that could have led to the positive culture.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.According to reviewing the instruction manual of hyf-xp, the reprocessing methods are described in the following chapters: chapter 6 compatible reprocessing methods and chemical agents.Chapter 7 cleaning, disinfection, and sterilization procedures.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES HYSTEROFIBERSCOPE
Type of Device
HYSTEROFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17690958
MDR Text Key322740257
Report Number3002808148-2023-09440
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04953170340178
UDI-Public04953170340178
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K891451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHYF-XP
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/23/2023
Initial Date FDA Received09/06/2023
Supplement Dates Manufacturer Received09/05/2023
11/13/2023
Supplement Dates FDA Received09/26/2023
11/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/11/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-