ADVANCED STERILIZATION PRODUCTS STERRAD® 100NX ALLCLEAR STERILIZER 1-DR DUO; STERRAD® 100NX EQUIPMENT Back to Search Results
Catalog Number 10104-007
Device Problems Device Emits Odor (1425); Environmental Particulates (2930)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/07/2023
Event Type  malfunction  
Event Description
A customer reported an event of a ¿smoke¿ or mist/haze and an ¿odor¿ or smell emitting from the sterrad® 100nx sterilizer.There was no report of any injuries or human reactions.The customer was instructed to power off the unit and discontinue use.An asp field service engineer was dispatched to assess the unit onsite.This event is being reported as a malfunction report subsequent to a previous serious injury.
 
Manufacturer Narrative
A field service engineer was dispatched to the customer site.The adapter converter, catalytic converter, oil mist filter, and vacuum pump oil were replaced to resolve the mist/haze and odor/smells issues.Unit meets specifications and was returned to service.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by advanced sterilization products, or its employees that the report constitutes an admission that the product, advanced sterilization products, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Asp complaint ref #: (b)(4).
 
Manufacturer Narrative
H3: asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the mist haze and odor/smells issues, and system risk analysis (sra).The device history record (dhr) was reviewed and no issues relating to the failure mode were noted.The involved unit met manufacturer specifications at the time of release.Trending analysis of the mist/haze and odor/smells issues for the sterrad® 100nx unit was reviewed for the prior six months from the event date, and no significant trend was observed.Review of risk documentation shows the risk for exposure to toxic or corrosive material to be "low." no parts were available for further analysis.The assignable cause of the mist/haze and odor/smell is likely due to the adapter converter, catalytic converter, oil mist filter, and vacuum pump oil.The asp field service engineer replaced the parts and confirmed the sterrad® 100nx met functional specifications after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD® 100NX ALLCLEAR STERILIZER 1-DR DUO
Type of Device
STERRAD® 100NX EQUIPMENT
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer (Section G)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer Contact
mark ongkeko
33 technology drive
irvine, CA 92618
9495619452
MDR Report Key17694423
MDR Text Key322777515
Report Number2084725-2023-00238
Device Sequence Number1
Product Code MLR
UDI-Device Identifier10705037048694
UDI-Public10705037048694
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10104-007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/09/2023
Initial Date FDA Received09/06/2023
Supplement Dates Manufacturer Received10/18/2023
Supplement Dates FDA Received10/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/26/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-