ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 14687-15 |
Device Problem
Material Rupture (1546)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/24/2023 |
Event Type
malfunction
|
Event Description
|
The event involved primary plum set, clave secondary port, clave y-site, secure lock, 103 inch in which the customer reported that the connector broke.There is unknown patient involvement; harm was no reported as a consequence of this event.
|
|
Manufacturer Narrative
|
The device is not available for investigation as the customer has discarded it.Without the return of the device, a probable cause is unable to be determined.
|
|
Manufacturer Narrative
|
A photo was provided showing the tubing separated from the y-clave.No samples were returned for investigation.The complaint of separation on the 14687-15 primary plum set can be confirmed from the photo provided.The probable cause is unknown.Without the return of the used sample a comprehensive failure investigation cannot be performed and a cause cannot be determined.The device history review for lot 13518721 was reviewed and no non conformities were found that would have led to the reported complaint.
|
|