ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0706-XTW
Device Problems Entrapment of Device (1212); Product Quality Problem (1506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/17/2023
Event Type  Injury  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, the reported product quality problem (irregular appearance) associated with the gripper sticking out of the clip arm after deployment was due to the gripper being unintentionally entangled with the leaflet and not being freed before deployment.The cause of the reported entrapment of device associated with the gripper entanglement could not be determined.The reported image resolution poor was associated with difficult visualization.The reported unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event Description
This is filed to report device entrapment.It was reported this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.An xtw clip was inserted, but while advancing under the valve, the anterior gripper became caught on the leaflet and was unable to be freed.Although unable to be removed from the leaflet, the clip was able to be deployed on both leaflets.It was noted after deployment, the gripper was sticking up outside the clip arms.One additional clip was implanted, reducing mr to a grade of 1.There was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key17695693
MDR Text Key322798264
Report Number2135147-2023-03884
Device Sequence Number1
Product Code NKM
UDI-Device Identifier05415067037381
UDI-Public05415067037381
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/23/2024
Device Catalogue NumberCDS0706-XTW
Device Lot Number30524R1103
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/17/2023
Initial Date FDA Received09/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
Patient Outcome(s) Required Intervention;
Patient Age85 YR
Patient SexFemale
Patient Weight65 KG
-
-