Model Number 799-EMVP-04-08 |
Device Problems
Communication or Transmission Problem (2896); Failure of Device to Self-Test (2937)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device displayed "internal comm failure -1474" and "off set self check failure - 1174" error messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.The device was put through extensive testing without duplicating the report.The smart pneumatic module board was replaced as a pre-caution.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|