Catalog Number 00663001600
Device Problem Patient Device Interaction Problem (4001)
Patient Problems Ossification (1428); Osteolysis (2377)
Event Date 06/15/2004
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2023-02431.G2: canada.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Event Description
It was reported a patient underwent a hip revision approximately eight years post implantation due to wear and osteolysis.During the revision heterotopic ossification was noted at the greater trochanter.The head and liner were replaced without complications.Attempts have been made and no further information is available.
 
Event Description
No additional event information to report at this time.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4; b5; d4; g3; h2; h4; h6 proposed component code: mechanical (g04)- stem no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: an initial right tha was performed with no complications noted.A mix of competitor and zb products were placed.A revision occurred approximately eight years later due to wear and osteolysis.During the revision, heterotopic ossification was noted at the greater trochanter.The zb head and liner were explanted and replaced with a mix of zb and competitor products.The complaint is confirmed based on the provided medical records.Dhr was not reviewed for the stem as the reported event was determined to be unrelated to the device.No problem was found, after a review by a health care professional it was determined that the heterotrophic ossification is unrelated to the stem.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL STEM 16MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17701815
MDR Text Key322877630
Report Number0001822565-2023-02430
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K902436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2001
Device Catalogue Number00663001600
Device Lot Number73950600
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/17/2023
Initial Date FDA Received09/07/2023
Supplement Dates Manufacturer Received10/23/2023
Supplement Dates FDA Received10/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/12/1991
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN LINER.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age75 YR
Patient SexMale
-
-