Model Number PVK-100S
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2023
Event Type  malfunction  
Event Description
The time of event is unknown.There was no report of patient harm.Video image failure(blackout).
 
Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb blackout.Based on the result, we concluded that it was caused due to the physical damage applied on the ccd drive pcb.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO ADAPTER MODULE EPK W/OUT EYEPC
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key17708522
MDR Text Key323100587
Report Number9610877-2023-58504
Device Sequence Number1
Product Code FEA
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPVK-100S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/23/2023
Initial Date FDA Received09/08/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-