• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC HEMOPRO POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC HEMOPRO POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED Back to Search Results
Model Number C-VH-3010
Device Problem Failure to Deliver Energy (1211)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2023
Event Type  malfunction  
Manufacturer Narrative
Tw id# (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, t.W.Power supply generator was not delivering energy.Everything was working normal when the hemopro suddenly stopped working.It was a sudden stop and not a gradual decrease in energy.They performed a series of troubleshooting and diagnosing scenarios by changing adapters and cords etc, and determined the power supply was the issue by hooking up a different power supply and it worked.Short delay while trouble shooting.No patient harm.
 
Manufacturer Narrative
Trackwise#: (b)(4).The reported device is an oem device.The certificate of conformance was reviewed for the serial # (b)(6).The vendor certifies that this device lot conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOPRO POWER SUPPLY
Type of Device
UNIT, CAUTERY, THERMAL, AC-POWERED
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key17711134
MDR Text Key322960361
Report Number2242352-2023-00763
Device Sequence Number1
Product Code HQO
UDI-Device Identifier00607567700826
UDI-Public00607567700826
Combination Product (y/n)N
PMA/PMN Number
K052274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC-VH-3010
Device Catalogue NumberC-VH-3010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/18/2023
Initial Date FDA Received09/08/2023
Supplement Dates Manufacturer Received10/05/2023
Supplement Dates FDA Received10/05/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/11/2023
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
ADAPTORS AND CORDS
-
-