• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND VELYS SAW HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND VELYS SAW HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 451570102
Device Problem Intermittent Loss of Power (4016)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 06/12/2023
Event Type  malfunction  
Event Description
This is report 2 of 2 for the same event.It was reported that during a total knee arthroplasty surgical procedure, it was observed that while using the robotic-assisted solution satellite station device and the robotic assisted saw handpiece device, it was observed that the saw stuttered.It was reported that the cuts were "continually inaccurate".Multiple attempts were made to gain additional information, and no additional information was provided.It was reported that all saw interface connections were check and were satisfactory.It was reported that the device was being used with a robotic assisted base station device.There were no delays in the procedure.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was returned for evaluation.During evaluation it was determined that the reported condition of saw stuttered could be confirmed.It was found that when the device was run, the blade tip appeared to oscillate up and down (out of plane) while the motor was driven.It is unknown when this failure occurred.Review of the device log files for this event determined that inaccurate cuts "continually" could not be confirmed but one case was confirmed as inaccurate possibly related to the saw.The most probable cause of the saw issue is loss of performance.(premature wear).No defects were found with the system or software.The overall assignable root cause was determined to be due to user/improper handling and component failure and premature wear.D10, concomitant medical devices and therapy dates, base station device, robotic assisted satellite station device.(b)(6), 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELYS SAW HANDPIECE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
DEPUY IRELAND
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthpaedic dr.
warsaw IN 46581
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17711836
MDR Text Key322973571
Report Number1818910-2023-18356
Device Sequence Number1
Product Code OLO
UDI-Device Identifier10603295519430
UDI-Public10603295519430
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number451570102
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2023
Initial Date Manufacturer Received 08/14/2023
Initial Date FDA Received09/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-