MAUDE Adverse Event Report: AMO MANUFACTURING USA, LLC VERITAS ADVANCED INFUSION PACK; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING USA, LLC VERITAS ADVANCED INFUSION PACK; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number VRT-AI
Device Problem Failure to Prime (1492)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that customer vrt-ai pack had priming issues (failed prime with error 503).Procedure was completed successfully using another pack.There was no patient contact.Product was returned for investigation and results confirmed excessive weld gap.It was also noted that pump tubing at the top of the cassette was slightly twisted contacting the top surface of the cassette.No further information provided.
 
Manufacturer Narrative
Additional narrative: note: this mdr report is being submitted due to a retrospective review of complaints associated with a remedial action (recall id number: 91290).Section a2, a3, a4, a5: unknown/ not provided.Section d6a/d6b - not applicable, as this is not an implantable device.Section b3: date of event: unknown, not provided, but the best estimate date is between (b)(6) 2022 and (b)(6) 2022.Section h9: johnson & johnson surgical vision (jjsv) has initiated a field action related to the veritas¿ advanced infusion packs (vrt-ai) and veritas¿ advanced fluidics packs (vrt-af) due to the potential for the weld protrusion to be larger than the design specification which could lead to failure during the priming cycle and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during the surgical case.This could result in a delay in surgery and/or longer surgical time, which may result in post-operative ocular sequalae, such as transient corneal edema.Device evaluation: product was returned and evaluated.A visual inspection showed excessive weld gap.The reported event was confirmed.Based on investigation results a product malfunction and product deficiency was confirmed.A nonconformance was initiated to address the weld gap protrusion findings.In addition the peristaltic pump tubing at the top of the cassette was slightly twisted and contacting the top surface of the cassette.The lower drain pump tubing was in contact with the bottom surface of the cassette.Manufacturing record review: a review of the records related to the device was performed.The records showed the device and its components met all specifications prior to being released.Conclusion: as a result of the investigation there is indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERITAS ADVANCED INFUSION PACK
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key17716002
MDR Text Key323049124
Report Number3012236936-2023-02245
Device Sequence Number1
Product Code HQC
UDI-Device Identifier35050474700902
UDI-Public(01)35050474700902(17)250927(10)60400378
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203060
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial
Report Date 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVRT-AI
Device Catalogue NumberVRT-AI
Device Lot Number60400378
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/16/2023
Initial Date FDA Received09/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2247-2023
Patient Sequence Number1
-
-