• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD Back to Search Results
Model Number 3186
Device Problem Low impedance (2285)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/17/2023
Event Type  Injury  
Event Description
It was reported that the patient's ipg is unable to enter mri mode due to low impedances.Troubleshooting was unable to resolve the issue.As a result, surgical intervention may be undertaken at a later date to address the issue.It is unknown which lead has the low impedances.
 
Manufacturer Narrative
Date of event is estimated.The allegation is against 1 of 2 leads; however, it is unknown which lead, therefore, all potential components are being listed.Additional components potentially involved in the event include: common device name: scs octrode lead, model: 3186, udi: (b)(4), serial: n/a, batch: 3126663.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Additional information indicates that surgical intervention was undertaken on (b)(6) 2024 wherein the ipg was electively explanted and replaced to address the issue.No intervention occurred on the leads.Effective therapy was restored post-op.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key17716461
MDR Text Key323036141
Report Number1627487-2023-04286
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05414734406130
UDI-Public05414734406130
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/28/2012
Device Model Number3186
Device Lot Number3126663
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/21/2023
Initial Date FDA Received09/08/2023
Supplement Dates Manufacturer Received11/01/2023
01/30/2024
04/06/2024
Supplement Dates FDA Received11/09/2023
02/22/2024
04/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS ANCHOR (2).; SCS IPG.; SCS LEAD.
Patient Outcome(s) Other;
Patient SexFemale
Patient Weight113 KG
-
-