• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL LIVEWIRE¿ TC ABLATION CATHETER, 4 MM TIP UNIVERSAL TEMPERATURE MONITORING, 115 C; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL LIVEWIRE¿ TC ABLATION CATHETER, 4 MM TIP UNIVERSAL TEMPERATURE MONITORING, 115 C; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number 402277
Device Problems Signal Artifact/Noise (1036); Use of Device Problem (1670); Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/17/2023
Event Type  malfunction  
Event Description
During the supraventricular tachycardia procedure, an expired catheter was used.It was noted there was a deflection issue of the catheter as well.When the catheter was inserted into the femoral vein, the issue was recognized on signal.Noise was detected on the electrode.The cable was replaced, with no resolution.The catheter was replaced, and the procedure was completed with no adverse consequences to the patient.
 
Event Description
During the paroxysmal supraventricular tachycardia procedure, a deflection issue was noted with the catheter.When the catheter was inserted into the femoral vein, the issue was recognized and noise was detected on the electrode.The cable was replaced, with no resolution.The catheter was replaced, and the procedure was completed with no adverse consequences to the patient.Additional information revealed this device was expired at the time of use.
 
Manufacturer Narrative
One 7f, quadripolar, medium curl, xls livewire tc catheter was received for evaluation.Functional testing revealed the catheter shaft was able to deflect when actuating the steering mechanism; however, the catheter no longer met specifications for curve shape in one direction.The catheter handle was opened, and the device was deflected.Once deflected, it was noted that the slide block was restricting deflection from meeting specifications.Electrodes 1-4 met specifications of acceptable resistance values with no open or short circuits detected.It should be noted that livewire tc ablation catheter batch number 7272743 expired on 31dec2022 which was prior to the reported event date of 17apr2023.A review of distribution records confirmed that this product was distributed prior to its expiration date.The cause of the reported noise remains unknown.Abbott is continuing to monitor the pull wire issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIVEWIRE¿ TC ABLATION CATHETER, 4 MM TIP UNIVERSAL TEMPERATURE MONITORING, 115 C
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345
Manufacturer (Section G)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key17726564
MDR Text Key323162899
Report Number2182269-2023-00041
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
P960016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model Number402277
Device Lot Number7272743
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/21/2023
Initial Date FDA Received09/11/2023
Supplement Dates Manufacturer Received11/03/2023
Supplement Dates FDA Received11/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/06/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-