Catalog Number 8065977763 |
Device Problems
Unintended Ejection (1234); Difficult to Advance (2920)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 08/15/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A facility representative reported with a description of intraocular lens (iol) stuck to plunger tip, unable to detach, lens was forcefully pushed to dislodge.There was patient contact.Additional information has been requested.
|
|
Manufacturer Narrative
|
The used company cartridge was returned with the opened pouch.Inadequate viscoelastic was observed in the cartridge.No damage was observed.The cartridge had evidence of placement into a cartridge.The cartridge was cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.A qualified lens model/diopter was indicated.The handpiece and viscoelastic information was not provided.It is unknown if qualified products were used.No problem was found with the returned cartridge.Top coat dye stain testing was conducted with acceptable results.The root cause for the reported advancement issue may be related to a failure to follow the instruction for use (ifu).There did not appear to be adequate viscoelastic in the cartridge.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.It is unknown if a qualified handpiece and viscoelastic were used.The ifu instructs: the company iol delivery system is for implantation of qualified company foldable iols.No unqualified lenses should be used with the company iol delivery system.The company cartridges are qualified for use with compatible company handpieces for the surgical implantation of company qualified foldable iols.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
Additional information has been received stating lens touch the cornea of the eye, but did not fully inserted into eye.
|
|
Manufacturer Narrative
|
The manufacturer internal reference number is: (b)(4).
|
|