Model Number A22002A |
Device Problems
Break (1069); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/22/2023 |
Event Type
malfunction
|
Event Description
|
A user facility reported to olympus the welding of the light guide connection part fell off and was buried in the light guide of telescope, 30°, 4 mm.There were no reports of patient or user harm associated with this event.
|
|
Manufacturer Narrative
|
The device was returned to an olympus repair facility, and an evaluation of the device was performed.During the evaluation, the customer's originally reported issue of the welding of the light guide connection part that fell off was reproduced.The following were also noted: bending of the outer tube, poor lighting due to a broken light guide, foreign matter adhered to the cover glass, and foreign matter adhesion was on the internal lens.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
|
Manufacturer Narrative
|
Correction: b3 and b5 were inadvertently omitted from the initial report.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The event likely occurred due to user error, improper handling as well as application of excessive force.Olympus will continue to monitor field performance for this device.
|
|
Event Description
|
It was reported that the event occurred following a therapeutic transcervical resection procedure.
|
|