Model Number G5A-80A |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
Complainant alleged that during functional testing, the device's defib output was out of specification.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of device data logs.However, the device was put through extensive testing including stress testing and full functional testing without duplicating the report.The main board and capacitors were replaced as a precaution.The device was recertified and returned for the device.The electrode pads were not returned as part of this investigation.Analysis of reports of this type has not identified an increase in trend.
|
|
Search Alerts/Recalls
|