An event of advancement difficulty through patient anatomy and catheter damage was reported.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Information from the field indicated that damage was observed at the opening of the catheter.The field also indicated that the patient had a high degree of calcareous plaque in the femoral artery.The device was returned for analysis.The nosecone and valve capsule were observed to be damaged/deformed.As the event is not reportable, a letter is not requested and there is no indication of a product issue, no device history record or lot specific similar complaint review is required.Based on available information, the cause of the reported event appears to be a result of calcareous plaque in the femoral artery.There is no indication of a product quality issue with respect to manufacture, design, or labeling.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
It was reported that on 13 january 2023, an unknown valve was chosen for implant using a large flexnav delivery system.It was noted that there was calcareous plaque in the femoral artery.While it was attempted to pass the delivery system through the femoral artery, resistance was encountered.The delivery system was removed, and damage was seen at the opening of the catheter.A new large flexnav delivery system was chosen to implant the valve.The valve was successfully positioned and implanted.The patient remained stable throughout the procedure and there was no clinically significant delay.The patient was reported as stable.
|