Model Number URF-V3
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation, and the customer's allegation was confirmed.The device evaluation found a leak from the bending section, a leak from the distal end, an objective lens discolored, a light guide lens discolored, corrosion inside the control section, the bending section was deformed, and a low angle issue.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The olympus field service engineer reported (on behalf of the customer) that the uretero-reno videoscope bending rubber was damaged.The issue was found during preparation for use before a diagnostic procedure.There were no reports of patient harm.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.It is likely that the bending part may have been damaged due to physical stress being applied to it.If handled according to the instructions for use (ifu) below, the user may be able to detect the event.Instruction manual urf-v3/v3r operation manual chapter 3 preparation and inspection users may be able to reduce/prevent the occurrence of this event by handling it in accordance with the ifu below.Instruction manual urf-v3/v3r operation manual - important information ¿ please read before use_ precautions:do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.- chapter 4 operation 4.1 warnings and cautions: operation :do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.:do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17733581
MDR Text Key323259704
Report Number9610595-2023-13237
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170435119
UDI-Public04953170435119
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/12/2023
Supplement Dates Manufacturer Received09/28/2023
Supplement Dates FDA Received10/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-