H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.H10: d4 (expiry date: 09/2013).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.Approximately six years and seven months post filter deployment, the patient had low back pain, x-ray of lumbar spine showed the filter with fracture of at least two and possibly three of its legs.Impression given that these fractured legs appeared to have mildly migrated a short distance away from the main portion of the filter; however, their exact location was indeterminate.Around one year, three months and three weeks later, x-ray of the upper gastrointestinal tract showed the filter with the tip at the l1-2 level.There were at least two fractured struts/legs of the filter noted.Therefore, the investigation is confirmed for filter limb detachment.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 09/2013), g3, h6 (method).H11: h6 (result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|