BOSTON SCIENTIFIC CORPORATION SPYGLASS DS; LED LIGHT SOURCE Back to Search Results
Model Number M00546650
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/21/2023
Event Type  malfunction  
Manufacturer Narrative
Block h6 (impact codes): imdrf impact code f1001 is being used to capture the reportable event of aborted/cancelled procedure.
 
Event Description
Note: this report pertains to a spyscope ds ii and spy ds controller used during the same procedure.It was reported to boston scientific corporation that a spyscope ds ii access & delivery catheter and spyglass digital controller were used in the biliary during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2023.During the procedure, when the spyscope ds ii was connected into the controller, no image was detected.Three dots came up then back to boston scientific screen.The procedure was not completed due to this event.
 
Manufacturer Narrative
Block h6 (impact codes): imdrf impact code f1001 is being used to capture the reportable event of aborted/cancelled procedure.Block h10 the returned spyglss controller was analyzed, as per sfmd report, normal wear and tear to top cover and front panel.Catheter interface contacts have unknown contamination and require replacement to repair this controller.The complaint was confirmed.A risk review confirms this is not a new or unanticipated event.As per report attached, the problem is unlikely related to manufacturing, as product analysis identified problem related to use and maintenance and did not identify any manufacturing defect.In addition, during a device history record review conducted by enercon, it was confirmed that the device met all manufacturing specifications.Although the number of procedures/recycles of the unit are unknown, improper handling of the device or wear/tear on internal components over time likely contributed to the event.Based on all gathered information, the conclusion code selected for this event is cause traced to maintenance, which indicates that problems are traced to improper routine or preventative maintenance.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event Description
Note: this report pertains to a spyscope ds ii and spy ds controller used during the same procedure.It was reported to boston scientific corporation that a spyscope ds ii access & delivery catheter and spyglass digital controller were used in the biliary during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2023.During the procedure, when the spyscope ds ii was connected into the controller, no image was detected.Three dots came up then back to boston scientific screen.The procedure was not completed due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPYGLASS DS
Type of Device
LED LIGHT SOURCE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
ENERCON TECHNOLOGIES
25 northbrook dr
665
gray ME 04039
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17736586
MDR Text Key323284576
Report Number3005099803-2023-04833
Device Sequence Number1
Product Code NTN
UDI-Device Identifier08714729874348
UDI-Public08714729874348
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00546650
Device Catalogue Number4665
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/21/2023
Initial Date FDA Received09/13/2023
Supplement Dates Manufacturer Received09/20/2023
Supplement Dates FDA Received10/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-