Note: this report pertains to a spyscope ds ii and spy ds controller used during the same procedure.It was reported to boston scientific corporation that a spyscope ds ii access & delivery catheter and spyglass digital controller were used in the biliary during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2023.During the procedure, when the spyscope ds ii was connected into the controller, no image was detected.Three dots came up then back to boston scientific screen.The procedure was not completed due to this event.
|
Block h6 (impact codes): imdrf impact code f1001 is being used to capture the reportable event of aborted/cancelled procedure.Block h10 the returned spyglss controller was analyzed, as per sfmd report, normal wear and tear to top cover and front panel.Catheter interface contacts have unknown contamination and require replacement to repair this controller.The complaint was confirmed.A risk review confirms this is not a new or unanticipated event.As per report attached, the problem is unlikely related to manufacturing, as product analysis identified problem related to use and maintenance and did not identify any manufacturing defect.In addition, during a device history record review conducted by enercon, it was confirmed that the device met all manufacturing specifications.Although the number of procedures/recycles of the unit are unknown, improper handling of the device or wear/tear on internal components over time likely contributed to the event.Based on all gathered information, the conclusion code selected for this event is cause traced to maintenance, which indicates that problems are traced to improper routine or preventative maintenance.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
Note: this report pertains to a spyscope ds ii and spy ds controller used during the same procedure.It was reported to boston scientific corporation that a spyscope ds ii access & delivery catheter and spyglass digital controller were used in the biliary during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2023.During the procedure, when the spyscope ds ii was connected into the controller, no image was detected.Three dots came up then back to boston scientific screen.The procedure was not completed due to this event.
|