BOLTON MEDICAL, INC. TREO ABDOMINAL STENT-GRAFT SYSTEM; STENT, ENDOVASCULAR GRAFT, AORTIC Back to Search Results
Catalog Number 28-L2-24-100S
Device Problems Break (1069); Retraction Problem (1536); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2023
Event Type  malfunction  
Manufacturer Narrative
Bolton medical, inc.(d/b/a terumo aortic), herein known as the "company", is submitting this report pursuant to 21 cfr part 803, has made reasonable efforts to obtain complete information, and has provided as much as is available to the company as of the submission date of this report.This report is based on information obtained by the company, which may not have been able to fully investigate or verify prior to the date the report was required by the regulations.Any required fields that are unpopulated are blank because the information is currently unknown, unavailable, or not applicable.This report does not constitute an admission or a conclusion by anyone that the device caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the certain regulations, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
"unable to deploy the leg.When rotating the gray knob the lead screw was coming back normally but the outer sheath was not retracting.The two components appeared to be disconnected.When the lever was exposed at the back end of the black stationary grip it appeared to be broken as per figure attached.When the leg was prepared, external box had no sign of damage and no maneuver were done that could have affect the integrity of the device." patient outcome - "no issue for the patient, procedure completed with back up device.".
 
Manufacturer Narrative
Section h6 health effect impact and medical device problem codes have been updated following completion of the investigation.Bolton medical, inc.(d/b/a terumo aortic), herein known as the "company", is submitting this report pursuant to 21 cfr part 803, has made reasonable efforts to obtain complete information, and has provided as much as is available to the company as of the submission date of this report.This report is based on information obtained by the company, which may not have been able to fully investigate or verify prior to the date the report was required by the regulations.Any required fields that are unpopulated are blank because the information is currently unknown, unavailable, or not applicable.This report does not constitute an admission or a conclusion by anyone that the device caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the certain regulations, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
"unable to deploy the leg.When rotating the gray knob the lead screw was coming back normally but the outer sheath was not retracting.The two components appeared to be disconnected.When the lever was exposed at the back end of the black stationary grip it appeared to be broken as per figure attached.When the leg was prepared, external box had no sign of damage and no maneuver were done that could have affect the integrity of the device." patient outcome - "no issue for the patient, procedure completed with back up device.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREO ABDOMINAL STENT-GRAFT SYSTEM
Type of Device
STENT, ENDOVASCULAR GRAFT, AORTIC
Manufacturer (Section D)
BOLTON MEDICAL, INC.
799 international parkway
sunrise FL 33325
Manufacturer Contact
megan indeglia
799 international parkway
sunrise, FL 33325
9548389699
MDR Report Key17741714
MDR Text Key323372465
Report Number2247858-2023-00242
Device Sequence Number1
Product Code MIH
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P190015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number28-L2-24-100S
Device Lot NumberB220518133
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/13/2023
Supplement Dates Manufacturer Received08/24/2023
Supplement Dates FDA Received12/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age83 YR
-
-