DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-45 |
Device Problems
Erratic or Intermittent Display (1182); Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) suspension salt water breaks and flows into the display screen causing the screen to become partially partitioned.There was no patient harm or injury reported.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Manufacturer Narrative
|
Updated data: b4, e1(initial reporter) e2, e3, g3, g6, h2, h10.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) had a issue of suspension salt water breaks and flows into the display screen.Patient demographics not available and no adverse event was reported.A getinge field service engineer inspected the unit and some areas of the display screen could not be displayed.There were traces of liquid intrusion.The display screen needed to be replaced, waiting for the customer to order accessories.On (b)(6) 2023, the display screen was replaced and the display function returned to normal.The equipment was tested according to factory requirements.All functional tests passed and can be delivered to customers.
|
|
Event Description
|
N/a.
|
|