Model Number 0998-00-0800-45
Device Problems Erratic or Intermittent Display (1182); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2023
Event Type  malfunction  
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) suspension salt water breaks and flows into the display screen causing the screen to become partially partitioned.There was no patient harm or injury reported.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
Updated data: b4, e1(initial reporter) e2, e3, g3, g6, h2, h10.
 
Event Description
N/a.
 
Manufacturer Narrative
It was reported that the cardiosave intra-aortic balloon pump (iabp) had a issue of suspension salt water breaks and flows into the display screen.Patient demographics not available and no adverse event was reported.A getinge field service engineer inspected the unit and some areas of the display screen could not be displayed.There were traces of liquid intrusion.The display screen needed to be replaced, waiting for the customer to order accessories.On (b)(6) 2023, the display screen was replaced and the display function returned to normal.The equipment was tested according to factory requirements.All functional tests passed and can be delivered to customers.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE I PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17745104
MDR Text Key323396201
Report Number2249723-2023-04037
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108421
UDI-Public10607567108421
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-45
Device Catalogue Number0998-00-0800-45
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/05/2023
Initial Date FDA Received09/14/2023
Supplement Dates Manufacturer Received09/15/2023
04/04/2024
Supplement Dates FDA Received09/19/2023
04/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/26/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN.
-
-