Model Number P1840DCA000001 |
Device Problem
Activation Problem (4042)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/21/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The hillrom technician found the cpr valve needed to be replaced.Per the hillrom service manual the totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Test the cpr release for correct operation and reset of the head cylinder.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the cpr valve to resolve the reported event.Based on this information, no further action is required.
|
|
Event Description
|
Hillrom received a report from a hillrom technician stating the bed cpr was inoperative.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
|
|