MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS DMC GMBH DIGITALDIAGNOST C90 HIGHPERFORMANCE; SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS DMC GMBH DIGITALDIAGNOST C90 HIGHPERFORMANCE; SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER) Back to Search Results
Model Number 712034
Device Problems Device Handling Problem (3265); Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported that a patient dropped a portable detector.A dropped detector can lead in worst case to a fracture in the foot.No injury reported.
 
Manufacturer Narrative
Ref.Id: (b)(4) the investigation is still ongoing on this event.When the investigation is completed a follow-up report will be sent to the fda.
 
Manufacturer Narrative
Ref.Id: (b)(4).The digitaldiagnost c90 is a stationary x-ray system for general radiographic purposes.As an option, a portable digital flat panel detector (model "skyplate") can be used for image capture.The local field service engineer (fse) went on site, checked the detector and tried several times to calibrate with no success.The affected detector need to be replaced.A new detector was sent to the site and was installed.Finally, the system meets the specification for the performed service and is returned to use.The customer confirms that the detector was dropped by accident (use error).Risk estimation revealed acceptable risk per risk benefit analysis, because actually, there is no feasible technical solution for this kind of ¿use error¿.This issue is further monitored and trended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIGITALDIAGNOST C90 HIGHPERFORMANCE
Type of Device
SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS DMC GMBH
roentgenstrasse 24
hamburg 22335
GM  22335
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS DMC GMBH
roentgenstrasse 24
hamburg 22335
GM   22335
Manufacturer Contact
dusty leppert
100 park avenue, suite 300
orange village, OH 44122
MDR Report Key17753574
MDR Text Key323750620
Report Number3003768251-2023-00050
Device Sequence Number1
Product Code MQB
UDI-Device Identifier00884838090699
UDI-Public00884838090699
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K182973
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number712034
Device Catalogue Number712034
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/30/2023
Initial Date FDA Received09/15/2023
Supplement Dates Manufacturer Received08/30/2023
Supplement Dates FDA Received11/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/20/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-