• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Catalog Number AB2000
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/28/2023
Event Type  malfunction  
Manufacturer Narrative
Component code 4756: per the instructions for use, the aquabeam motorpack, a re-usable component of the aquabeam robotic system, provides power to the aquabeam handpiece by means of dc motors root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation (procept) became aware that during the aquablation procedural set-up, there was no connection between the aquabeam motorpack and the aquabeam console.As a result, the aquablation procedure was aborted and converted to transurethral resection of the prostate (turp) surgery.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
The aquabeam motorpack was returned for investigation.During functional testing, an attempt was made to jog the motors on the aquabeam motorpack and was unsuccessful.The aquabeam motorpack was disassembled and the cable subassembly was replaced with a known good one and connected to the test system.The cpu could detect the motorpack connection, turning all indicators to green.The motors responded by jogging.The root cause of the reported error is a malfunctioning cable; however, the cause of the malfunctioning cable is undeterminable.A review of the device history record (dhr) ab2000-b / serial number (b)(6) and aquabeam motorpack / lot number 21c01779 was conducted, which confirmed there were no non-conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system met all design and manufacturing specificiations when released for distribution.The current user manual um0104-00 rev.G, aquabeam robotic system user manual, intl (ce), english was reviewed.11.2.3 non-sterile: console, motorpack, and foot pedal connection connect the motorpack cable to the front of the console: connect the motorpack plug on the front right port.Ensure the connector locks are in place and the red marks on the motorpack and the console align submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr
san jose CA 95134
Manufacturer Contact
doria esquivel
150 baytech dr
san jose, CA 95134
6502327291
MDR Report Key17754510
MDR Text Key323475482
Report Number3012977056-2023-00150
Device Sequence Number1
Product Code PZP
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/29/2023
Initial Date FDA Received09/15/2023
Supplement Dates Manufacturer Received11/30/3023
Supplement Dates FDA Received12/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/29/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-