MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Catalog Number AB2000
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/29/2023
Event Type  malfunction  
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation (procept) became aware that during the aquablation procedure, the aquabeam handpiece was noted to be broken.A second handpiece was opened to complete the procedure.The reported event caused a surgical procedure delay of over 20 minutes.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
Root cause of the reported event has not yet been established.Investigation by manufacturer is currently in process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Manufacturer Narrative
The aquabeam handpiece was returned for investigation.Visual inspection of the returned device confirmed the reported broken handpiece, as the aspiration tubing and irrigation tubing were observed to be detached from the manifold.Functional testing could not be performed as the stainless steel tubing assembly was pushed out.Thus, how the handpiece broke while translating the scope remains undeterminable.Likely, the scope was not latched properly, which led to the incorrect positioning of the jet probe.The root cause is undeterminable.A review of the device history record (dhr) (b)(4)/ serial number (b)(6) and aquabeam handpiece / lot number 23c00294 was conducted, which confirmed that there were two (2) non-conformance issued to this lot during the manufacturing process that could potentially be related to the reported event.The affected units within the lot were reworked to address the nonconformance.Upon re-inspection, the lot met all required specifications and was then deemed acceptable to be released for distribution per device specifications.The current user manual um0101-00 rev.F, aquabeam robotic system user manual, us, english was reviewed.11.2.5 sterile: aquabeam handpiece and aquabeam scope setup hold the distal end of the scope tube tip approximately 1 inch (2.54 cm) from the fully proximal position and continue advancing the aquabeam scope forward until it is properly engaged with the aquabeam handpiece and then rotate the proximal key alignment adapter so that the dimple on the proximal key alignment adapter is facing up.An audible click should be heard when the aquabeam scope is securely engaged with the aquabeam handpiece.Note: do not use excessive force to advance the aquabeam scope.If resistance is felt,gently rotate the aquabeam scope clockwise and counterclockwise while simultaneously applying forward pressure.Submission of this report does not constitue an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
san jose CA 95134
Manufacturer Contact
doria esquivel
150 baytech drive
san jose, CA 95134
6502327291
MDR Report Key17754986
MDR Text Key323492039
Report Number3012977056-2023-00151
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/16D20220318E
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/23/2023
Initial Date FDA Received09/15/2023
Supplement Dates Manufacturer Received11/17/2023
Supplement Dates FDA Received11/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-