LIVANOVA USA, INC. PULSE GEN MODEL 1000; GENERATOR Back to Search Results
Model Number 1000
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/22/2023
Event Type  malfunction  
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
It was reported that during a battery replacement, when interrogating the new replacement generator, error code 14 (generator disabled due to brown-out-reset) was observed.Per the report, the error code was seen when the generator was taken out of the package, and it was not suspected that the device was exposed to any electrocautery as the error code was seen before being implanted.After re-enabling therapy, when the device was re-interrogated, diagnostics were all ok and the generator was implanted.Tablet data was provided for review.No other relevant information has been received to date.
 
Event Description
Information was received from the surgeon that the device was stored inside its sterile package and that the storage temperature prior to surgery was "body temperature" (i.E.Normal).
 
Event Description
Additional information was received reporting that the patient was recently seen by their neurologist and diagnostics were within normal limits.No additional report of resets.No other relevant information has been received to date.
 
Manufacturer Narrative
This report was due on november 25th, 2023; however, due to a network disruption at livanova, the ability to submit mdrs was lost on november 20, 2023, before this report was ready to submit.The report was prepared and submitted immediately following restoration of the livanova systems.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 1000
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key17756984
MDR Text Key323592055
Report Number1644487-2023-01333
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750405
UDI-Public05425025750405
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1000
Device Lot Number7620
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 08/22/2023
Initial Date FDA Received09/15/2023
Supplement Dates Manufacturer Received09/20/2023
10/26/2023
Supplement Dates FDA Received10/13/2023
12/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age9 YR
Patient SexMale
-
-