• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TENEX HEALTH, INC. TX2 TISSUE REMOVAL SYSTEM - MICROTIP; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TENEX HEALTH, INC. TX2 TISSUE REMOVAL SYSTEM - MICROTIP; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number 554-2003-001
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/15/2023
Event Type  malfunction  
Event Description
During a procedure with the tenex system, the microtip needle separated from the rest of the hand piece.The procedure was completed with another microtip hand piece.There were no patient complications.
 
Manufacturer Narrative
Follow-up report #01.Sections b4, d9, g6, h2, h3, h6 and h10 have been updated with new information.The device was returned and evaluated.It was confirmed that the microtip needle had separated from the rest of the handpiece.Due to the state in which the device was received, functional testing could not be performed.The fracture site did not immediately indicate a specific cause for the failure.Disassembly of the device did not reveal any further anomalies or defects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TX2 TISSUE REMOVAL SYSTEM - MICROTIP
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
TENEX HEALTH, INC.
26902 vista terrace
lake forest CA 92630 8123
Manufacturer (Section G)
TENEX HEALTH, INC.
26902 vista terrace
lake forest CA 92630
Manufacturer Contact
david vancelette
26902 vista terrace
lake forest, CA 92630
9492388220
MDR Report Key17760963
MDR Text Key324123445
Report Number1000135560-2023-00036
Device Sequence Number1
Product Code LFL
UDI-Device Identifier00857156006154
UDI-Public(01)00857156006154(10)0552302(17)250223
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153299
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number554-2003-001
Device Catalogue Number554-2003-001
Device Lot Number0552302
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 08/16/2023
Initial Date FDA Received09/15/2023
Supplement Dates Manufacturer Received08/16/2023
Supplement Dates FDA Received10/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-