HOLOGIC, INC BREVERA BREAST BIOPSY SYSTEM; INSTRUMENT, BIOPSY Back to Search Results
Model Number BREVDISP09
Device Problem Failure to Obtain Sample (2533)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2023
Event Type  Injury  
Event Description
It was reported that during a brevera procedure on (b)(6) 2023, the needle did not fire during the biopsy after passing al quality checks.A second needle was used to complete the procedure.Due to this the patient required a second incision on the breast.No additional harm to the patient was reported.No other information is available.
 
Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BREVERA BREAST BIOPSY SYSTEM
Type of Device
INSTRUMENT, BIOPSY
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
daniel guevara
562 parkway
coyol free zone building b24
san jose 20102
MDR Report Key17765346
MDR Text Key323620833
Report Number1222780-2023-00351
Device Sequence Number1
Product Code KNW
UDI-Device Identifier15420045512863
UDI-Public(01)15420045512863(17)250525(10)23E26RF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163052
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberBREVDISP09
Device Catalogue NumberBREVDISP09
Device Lot Number23E26RF
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 09/05/2023
Initial Date FDA Received09/18/2023
Supplement Dates Manufacturer Received09/05/2023
Supplement Dates FDA Received10/12/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-