• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD NEXIVA¿ CLOSED IV CATHETER SYSTEM; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD NEXIVA¿ CLOSED IV CATHETER SYSTEM; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383516
Device Problem Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2023
Event Type  malfunction  
Manufacturer Narrative
D.4.Medical device expiration date: unknown.H.3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.H.4.Device manufacture date: unknown.
 
Event Description
It was reported that the bd nexiva¿ closed iv catheter system tubing expanded when used with the power injector.The following information was provided by the initial reporter: "re cat 383516 nexiva - customer claims "bd nexiva used for ct scan with power injector.Psi was under pressure limit of 300psi.Db nexiva catheter dilatated from wings of iv to the two port hub."".
 
Manufacturer Narrative
Investigation summary bd was unable to perform a thorough investigation as no sample, lot, or batch number were provided.A device history review could not be completed as no batch number was provided.
 
Event Description
It was reported that the bd nexiva¿ closed iv catheter system tubing expanded when used with the power injector.The following information was provided by the initial reporter: "re cat 383516 nexiva - customer claims "bd nexiva used for ct scan with power injector.Psi was under pressure limit of 300psi.Db nexiva catheter dilatated from wings of iv to the two port hub."".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD NEXIVA¿ CLOSED IV CATHETER SYSTEM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
jennifer suh
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17770159
MDR Text Key323844316
Report Number1710034-2023-01073
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00382903835164
UDI-Public00382903835164
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K183399
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number383516
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/31/2023
Initial Date FDA Received09/18/2023
Supplement Dates Manufacturer Received10/17/2023
Supplement Dates FDA Received10/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-