A healthcare facility in france reported via a fisher and paykel healthcare (f&p) field representative that a pt101 airvo 2 humidifier generated a spark during patient use.It was noted that a black mark was observed on the power cord outlet of the device.There were no reported patient consequences.Further information regarding the reported event, including the sequence of events, was requested from the healthcare facility.
|
A healthcare facility in france reported via a fisher and paykel healthcare (f&p) field representative that a pt101 airvo 2 humidifier generated a spark during patient use, followed by a dim screen.As a result, there was a black mark on the power cord outlet of the device.There were no reported patient consequences.F&p requested further information about the event, however no response was provided.
|
(b)(4).Product background: the pt101 airvo 2 humidifier (airvo 2) device is an active humidifier with integrated flow generator that delivers high flow warmed and humidified respiratory gases to spontaneously breathing patients through a variety of patient interfaces.It's intended use is for the treatment of spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.The airvo 2 device should not be used for life support purposes, and appropriate patient monitoring must be used at all times.Method: the complaint airvo 2 device and power cord was received by fisher & paykel healthcare (f&p), where it was inspected by a trained f&p employee.Several attempts to request further information from the healthcare facility were made, including the sequence of events and device set-up.However, no response was provided.Our investigation is baed on the initial information provided by the healthcare facility, our evaluation of the subject device, previous investigations of similar complaintts, and our knowledge of the product.Results: visual inspection of the subject device's exterior revealed no signs of impact damage.However, visible marks were observed on the mains inlet socket cover of the device, confirming the reported event.Inspection of the subject power cord revealed that there was an open circuit along the positive wire in the lead.The subject airvo 2 was powered on using a known good power cord, and the airvo 2 device operated as intended, and had no issues with the device screen.Conclusion: we are unable to determine the cause of the reported event.However, based on previous investigations it is possible the damage to the internal wiring of the power cord may be caused by excessive bending force.Excessive bending force can cause the internal wiring to detach from the crimp, leading to a potential arcing (spark).All f&p airvo 2 humidifiers undergo inspection during the production.The reported airvo 2 device would have met the required specification at the time of production.Our user instructions which accompany the pt101 airvo 2 humidifier state the following: avoid unncessary removal of the power cord from the rear of the device.Avoid pulling on the power cord.Appropriate patient monitoring must be used at all times.
|