Catalog Number RF320J
Device Problems Patient-Device Incompatibility (2682); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/04/2020
Event Type  malfunction  
Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.H10: d4 (expiry date: 04/2011).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed for a patient with right lower extremity deep vein thrombosis and saddle pulmonary embolism.Approximately twelve years and three months post filter deployment, it was alleged that the filter struts had perforated and detached.The device has not been removed and there were no reported attempts made to retrieve the filter.There was no reported patient injury.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed for a patient with right lower extremity deep vein thrombosis and saddle pulmonary embolism.Approximately twelve years and three months post filter deployment, it was alleged that the filter struts had perforated and detached.The device has not been removed and there were no reported attempts made to retrieve the filter.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.Approximately twelve years and three months post filter deployment, computed tomography scan of abdomen without intravenous contrast showed multiple strut perforation.One strut possibly penetrated the aorta, and another appeared to penetrate a bowel loop.In addition, review of computed tomography examination revealed a possible bent/fractured strut.Therefore, the investigation is confirmed for the reported perforation of inferior vena cava and filter limb detachment.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiry date: 04/2011), g3, h6 (method).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G2 FILTER SYSTEM - JUGULAR
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17773392
MDR Text Key323725045
Report Number2020394-2023-00703
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer,Health Professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberRF320J
Device Lot NumberGFSD2361
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/23/2023
Initial Date FDA Received09/19/2023
Supplement Dates Manufacturer Received11/17/2023
Supplement Dates FDA Received11/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/30/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexFemale
-
-