• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS INTELLIVUE MX40 2.4GHZ

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS INTELLIVUE MX40 2.4GHZ Back to Search Results
Model Number 865351
Device Problem Inaudible or Unclear Audible Prompt/Feedback (2283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/28/2023
Event Type  malfunction  
Manufacturer Narrative
Philips is in process of obtaining additional information.A final report will be submitted upon completion of the investigation.
 
Event Description
During evaluation at bench repair, it was identified that the device had no audio.Patient involvement is unknown.There was no report of patient or user harm.
 
Manufacturer Narrative
Diagnostic/functional testing was performed at the philips authorized repair facility.Results of functional testing indicate that the speaker produced no sound and the speaker was defective.Based on the information available and the testing conducted, the cause of the reported problem was a defective speaker.The reported problem was confirmed.The customer was previously provided a replacement device to resolve the issue.It has been concluded that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX40 2.4GHZ
Type of Device
INTELLIVUE MX40 2.4GHZ
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key17773771
MDR Text Key323730374
Report Number1218950-2023-00661
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation 505
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number865351
Device Catalogue Number865351
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/28/2023
Initial Date FDA Received09/19/2023
Supplement Dates Manufacturer Received06/28/2023
Supplement Dates FDA Received10/06/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-