• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEART START XL; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEART START XL; DEFIBRILLATOR Back to Search Results
Model Number M4735A
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2023
Event Type  malfunction  
Event Description
Philips received a complaint on the heartstart xl indicating that the panel of the display is aging and can't see clearly.The customer was informed that service could no longer be completed on the unit as the device had reached end of life (eol).According to service bulletin sb86100166d, the m4735a heart/start xl portable defibrillator/monitor was discontinued on 31-december -2013.All options associated with this defibrillator were discontinued as of 31-december-2013.The end of support date for the device is 31-december-2018.The customer was aware of the end of life terms and the device remains at the customer site.As troubleshooting was not completed for the device, the reported issue could not be verified and a cause could not be established.The reported problem was not confirmed.Based on the information available and results of additional analysis, no further action is necessary at this time.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The customer was made aware of the end of life terms and the device remains at the customer site.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART START XL
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
tanya deschmidt
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key17774009
MDR Text Key323733486
Report Number3030677-2023-03712
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K021453
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM4735A
Device Catalogue NumberM4735A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/14/2023
Initial Date FDA Received09/19/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-