MAUDE Adverse Event Report: ABBOTT JOT DX¿ ICM; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT JOT DX¿ ICM; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION) Back to Search Results
Model Number DM4500
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/18/2023
Event Type  Injury  
Event Description
It was reported that the device could not be interrogated.The device was explanted and replaced to resolve the event.The patient was stable and there were no adverse consequences.
 
Manufacturer Narrative
The reported event of failure to interrogate could not be confirmed.The device was received in normal working conditions with battery voltage above elective replacement indicator (eri).Analysis performed, indicated normal device functionality.Interrogation was performed on both merlin 1 and 2 programmers and showed normal interrogation.Performed dhr review to ensure that each manufacturing and inspection operation was performed.The product passed all quality control tests prior to its distribution.Longevity assessment was performed and device was in the normal range of operation with appropriate remaining longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOT DX¿ ICM
Type of Device
RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17774326
MDR Text Key323739332
Report Number2017865-2023-46462
Device Sequence Number1
Product Code MXC
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDM4500
Device Lot NumberS000090260
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/18/2023
Initial Date FDA Received09/19/2023
Supplement Dates Manufacturer Received11/10/2023
Supplement Dates FDA Received11/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-