Catalog Number UNK HIP FEMORAL STEM |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Foreign Body Reaction (1868); Discomfort (2330)
|
Event Date 08/17/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint # (b)(4).D4: the device catalog number is unknown; therefore, udi is unavailable.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
It was reported that the patient was experiencing discomfort and masses were noted via imaging around a well fixed unknown prosthesis.On 8-17-23 surgeon debrided masses and when the head was removed, trunnion debris was present.Surgeon chose to replace head with a delta ts in hopes that the similar metals was reduce further debris generation.Doi: unknown.Dor: (b)(6) 2023.Affceted side: left hip.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).Investigation summary no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot the device lot number is unknown, therefore a device history review could not be performed. if the lot/serial number becomes available, the record will be re-assessed.
|
|
Search Alerts/Recalls
|
|